8600 Series VSM
Lake Shore
8600 Series VSM
More science, less time
Advanced measurement performance
The 8600 Series vibrating sample magnetometer raises the bar for magnetometer performance and convenience. These VSMs combine high sensitivity (15 nemu), rapid measurement speed (10 ms/pt), and simple operation for more accurate measurements, faster.
Convenient operation
The entire 8600 Series system has been reimagined with a focus on clean, ergonomic design that simplifies the researcher’s interaction with the system. A motorized head brings the sample to a comfortable height for easy, one-handed exchange of the QuickLIGN™ sample rods. Temperature options include a cryostat, high-temperature oven, and single stage variable temperature insert. The combined temperature range of the options is 4.2 K to 1273 K. All three GlideLOCK™ options quickly slide into place and are auto-detected, with the software automatically displaying controls for the specific option. The magnet poles are also easily adjusted with the ExactGAP™ indexed positioning system that allows the pole gap to be set at one of six repeatable positions, eliminating the need to recalibrate after each change.
Designed for FORC
This VSM was created with first order reversal curve (FORC) measurement as a primary objective. FORC analysis is greatly enhanced by the high sensitivity of the 8600 Series vibrating sample magnetometer. FORC also benefits from increased data point density, and the new VSM flies through complex FORC data collection sequences in a fraction of the time required on previous systems.
Download our FREE FORC measurements ebook
Flexible and adaptable
The system’s software simplifies control of the VSM. Standard predefined measurement routines are combined with configurable field and measurement loops to provide a flexible data acquisition environment. In addition, the software features an integrated scripting tool, which enables the user to extend the existing routines with an open-ended software scripting language that can be used to perform customized measurement protocols and interface with third-party lab equipment.
8600 Series Features
- 15 × 10-7 emu noise floor at 10 s/pt
- 10 ms/pt data acquisition rate
- 10,000 Oe/s field ramp rate
- Rapid, repeatable temperature option exchange
- High stability—±0.05% per day
- Fields to 3.26 T
- Widest available temperature range — 4.2 K to 1273 K
8600 Series VSM Specifications
APPLIED FIELD STRENGTH WITH HIGH FIELD POLE CAPS (±1%) |
|||||
Gap indices 4 and 6 support non-VSM applications |
Sample space |
Sample size with option |
8604 maximum field |
8607 maximum field |
8607 maximum field |
ExactGAP™ Index 1 |
3.5 mm (0.14 in) |
N/A |
27.6 kOe (2.76 T) |
32.6 kOe (3.26 T) |
24.0 kOe (2.4 T) |
ExactGAP™ Index 2 |
8 mm (0.31 in) |
N/A |
25.2 kOe (2.52 T) |
30.1 kOe (3.01 T) |
21.0 kOe (2.1 T) |
ExactGAP™ Index 3 |
16 mm (0.63 in) |
6.4 mm (0.25 in) |
20.3 kOe (2.03 T) |
26.2 kOe (2.62 T) |
19.0 kOe (1.9 T) |
ExactGAP™ Index 5 |
24 mm (0.94 in) |
6.4 mm (0.25 in) |
15.5 kOe (1.55 T) |
23.1 kOe (2.31 T) |
15.0 kOe (1.5 T) |
*ExactGap 1 or 2 maximum ramp rate = 5 kOe/s (0.5 T/s); ExactGap 3 maximum ramp rate = 4 kOe/s (0.4 T/s); ExactGap 5 maximum ramp rate = 3 kOe/s (0.3 T/s) |
APPLIED FIELD STRENGTH WITH STANDARD POLE CAPS (±1%) |
|||||
Gap indices 4 and 6 support non-VSM applications |
Sample space |
Sample size with option |
8604 maximum field |
8607 maximum field |
8607 maximum field |
ExactGAP™ Index 1 |
3.5 mm (0.14 in) |
N/A |
26.1 kOe (2.61 T) |
31.2 kOe (3.12 T) |
23.0 kOe (2.30 T) |
ExactGAP™ Index 2 |
8 mm (0.31 in) |
N/A |
23.7 kOe (2.37 T) |
28.9 kOe (2.89 T) |
20.3 kOe (2.03 T) |
ExactGAP™ Index 3 |
16 mm (0.63 in) |
6.4 mm (0.25 in) |
19.6 kOe (1.96 T) |
25.3 kOe (2.53 T) |
18.5 kOe (1.85 T) |
ExactGAP™ Index 5 |
24 mm (0.94 in) |
6.4 mm (0.25 in) |
15.4 kOe (1.54 T) |
22.3 kOe (2.23 T) |
14.5 kOe (1.45 T) |
*ExactGap 1 or 2 maximum ramp rate = 5 kOe/s (0.5 T/s); ExactGap 3 maximum ramp rate = 4 kOe/s (0.4 T/s); ExactGap 5 maximum ramp rate = 3 kOe/s (0.3 T/s) |
MOMENT MEASUREMENT |
|
Noise floor (emu RMS) Room temperature |
|
0.1 s/pt |
150 nemu |
1 s/pt |
60 nemu |
10 s/pt |
15 nemu |
With oven or cryostat option, 10 s/pt |
300 nemu |
With SSVT option 10 s/pt |
90 nemu |
With vector coil option |
x coils: 90 nemu at ExactGAP 3; y coils: 200 nemu at ExactGAP 3 |
Dynamic range |
25 × 10-9 to 103 emu |
Moment stability |
Better than ±0.05% of full scale/day for fixed coil geometry at constant field and temperature |
Reproducibility |
Better than ±0.5% of reading, or ±0.1% of full scale, fixed rotation angle and range, with sample replacement |
Moment accuracy |
Better than 1% of reading ±0.2% of full scale with a geometrically identical test sample and calibrant |
Sample mass |
0 to 10 g |
FIELD MEASUREMENT |
|
Field accuracy |
1% of reading or ±0.05% of full scale |
Field resolution |
1 mOe |
Field ramp rate |
8604: 10,000 Oe/s |
Closed loop field control stability |
1 mOe |
SAMPLE ROTATION |
|
Setting resolution |
0.1° |
Setting reproducibility |
<1° |
Rotation range |
Continuous |
CERTIFICATIONS |
|
CE |
Yes |
Application of Council directives |
2014/35/EU Low Voltage Directive |
Standard to which conformity is declared |
EN61010-1: 2010 Overvoltage Category II, Pollution Degree 2; |
UTILITIES |
|
Total system cooling water power dissipation (50 or 60 Hz) — see available recirculating chillers |
8604: 4250 W |
Note: chiller power specifications listed are for 100% duty cycle. Many common magnet testing applications such as hysteresis loops will not require as much cooling power. Please consult Lake Shore for chiller recommendations for your specific applications. |
8600 Series VSM equipment
|
8604 |
8607 |
VSM head drive |
Gen2 |
|
VSM frame |
V-mag |
|
Control electronics |
Model 737 |
|
Bipolar magnet power supply |
Model 643 |
Model 648 |
Maximum output |
±35 V/±70 A (2,450 W) |
±75 V/±135 A (9.1 kW nominal) |
AC line input |
200/208 VAC ±10%, 13 A/phase; |
200 VAC ±5%, 41 A/phase; |
Cooling water requirements |
Tap water or closed cooling system |
|
Flow rate |
5.7 L (1.5 gal)/min minimum |
7.6 L (2.0 gal)/min minimum |
Pressure drop |
10 kPa (1.5 psi) at 5.7 L (1.5 gal)/min |
159 kPa (23 psi) at 7.6 L (2 gal)/min |
Electromagnet |
Model EM-4V |
Model EM-7V |
Pole diameter |
100 mm (4 in) |
180 mm (7 in) |
Pole cap face diameter |
50 mm (2 in) |
50 mm (2 in) |
Field homogeneity |
±0.1% over 1 cm3 (0.4 in3) |
±0.1% over 1 cm3 (0.4 in3) |
Cooling water requirements |
Tap water or closed cooling system |
|
Inlet temperature |
15 to 25 °C (59 to 77 °F) |
15 to 32 °C (59 to 89 °F) |
Flow rate |
7.6 L (2 gal)/min |
11.4 L (3 gal)/min |
Pressure drop |
200 kPa (30 psi) |
220 kPa (32 psi) |
Water chiller capacity |
2.5 kW (8,530 BTU)/h |
5 kW (17,060 BTU)/h |
Hall probe |
86-0911 |
|
Sense coils |
Model 86-SC and Model 86-LC |
|
Instrument console |
483 mm (19 in) rack |
8600 Series VSM option specifications
Model 86-CRYO variable temperature cryostat |
||
With LHe |
Temp range |
4.2 K (base), 5.5 K to 450 K (control) |
Temp stability |
±0.1 K |
|
With LN2 |
Temp range |
77.6 K (base), 85 K to 450 K (control) |
Temp stability |
±0.2 K |
|
Temperature resolution |
0.001 K |
|
Cool-down time |
5 min (15 min initial cool-down) |
|
Operation method |
Continuous flow |
|
LHe liquid usage |
<1 L/h when operating >7 K |
|
LN2 liquid usage |
<1 L/h when operating >85 K |
|
Insulation |
Vacuum |
|
Sample zone dimensions |
Bore size |
7.1 mm (0.28 in) |
Outside diameter |
22.4 mm (0.88 in) |
More information about VSM options
Model 86-OVEN high temperature oven |
||
Temperature range |
308 K to 1273 K |
|
Temperature stability |
±0.1 K |
|
Temperature resolution |
0.001 K |
|
Nominal ramp rate |
5 K/min |
|
Insulation |
Vacuum plus multiple reflective shields |
|
Sample zone dimensions |
Bore size |
7.3 mm (0.29 in) |
Outside diameter |
19 mm (0.75 in) |
More information about VSM options
Model 86-SSVT single stage variable temperature option |
||
Temperature range |
78 K (base), 100 K to 950 K (control) |
|
Temperature stability |
0.1 K RMS |
|
Temperature resolution |
0.001 K |
|
Gasses |
LN2 and nitrogen gas for T<350 K; |
|
Cool-down time |
40 min from room temp to 100 K |
|
Nominal ramp rate |
Up to 8 K/min |
|
Operating method |
Continuous flow |
|
LN2 liquid usage |
0.5 L/h >100 K to 350 K |
|
Nitrogen gas usage |
~3 L/min 100 K to 350 K |
|
Argon gas usage |
~3 L/min |
|
Insulation |
Vacuum |
|
Sample zone dimensions |
Bore size |
7.8 mm (0.31 in) |
Outside diameter |
12.7 mm (0.5 in) |
More information about VSM options